- Only for use in conjunction with the Millimar C 1240 M compact length measuring instrument
- Maximum measuring accuracy and minimum linearity deviation <0.01 %, i.e. 0.4 ?m over the entire measuring range
- Probe protected against dirt and moisture, can therefore be used close to production
- Chemical resistance data: resistant to oil, gasoline, water and aliphatic compounds. Moderately resistant to acids, bases, solvents and ozone
- Technical data
Order no. | Measuring range from | Product type | Measuring range up to | Distance of lower stop MIN | Distance of lower stop MAX | Distance of upper stop MAX | Lifter / retraction | Measuring force (N) | Increase in measuring force | Sensitivity deviation | Repeatability fw (µm) | Hysteresis fu (µm) | Linearity deviation within +/- 1,0 mm | Linearity deviation within +/- 2,0 mm | IP protection category | Cable length | Working temperature | Working temperature MIN | Working temperature MAX | Operating temperature Min | Operating temperature Max | Storage temperature MIN | Storage temperature MAX | Temperature coefficient | Compatibility |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
5313400 | -2 mm | 1340 | 2 mm | -2.2 mm | 0 mm | 3 mm | Vacuum lifter | 0,75 N | 0,08 N/mm | 0.3 % | 0.08 µm | 0.08 µm | 0.15 µm | 0.4 µm | IP 64 | 1.5 m | 20 °C | 10 °C | 40 °C | -10 °C | 80 °C | -10 °C | 80 °C | 0.6 µm/°C | Mahr 1340 |